9 To Replace Load Cell Assembly1. Disassemble top housing of scale.2. Disconnect the load cell wires from main PCB. Remove the screws fixing thebottom plate of the load cell assembly. Replace the load cell with a new one.Connect the load cell wires and fix the screws.3. Perform internal calibration as described in section 4.4. Assemble the top housing.5. Check accuracy of scale at different weight.EMB 200-2 EMB 600-2 EMB 220-1 EMB 500-1weight(g)tol.(g)weight(g)tol.(g)weight(g)tol.(g)weight(g)tol.(g)50.00 0.02 100.00 0.02 50.0 0.2 100.0 0.3100.00 0.02 200.00 0.02 100.0 0.2 300.0 0.3150.00 0.02 400.00 0.04 150.0 0.2 500.0 0.3200.00 0.02 600.00 0.06 200.0 0.2EMB 1200-1 EMB 2200-0 EMB 5.2K1 EMB 5.2K5weight(g)tol.(g)weight(g)tol.(g)weight(g)tol.(g)weight(g)tol.(g)200.0 0.3 500 2 1000 3 1000 10500.0 0.3 1000 2 3000 3 3000 101000.0 0.3 1500 2 5000 3 5000 101200.0 0.3 2000 2EMB 100-3 EMB 1000-2weight(g)tol.(g)weight(g)tol.(g)10.000 0.002 100.00 0.0220.000 0.002 200.00 0.0250.000 0.003 500.00 0.0370.000 0.003 700.00 0.03100.000 0.003 1000.00 0.036. Check other functions, such as Tare and Auto-Off.EMB-SH-e-0933 11
PreviousNext |