EOB_M-BA-e-071016Position of adjustment switch:Position of adjustment switch Status1. to the left Not documented2. concentric adjustment position – enables adjustment3. to the right Verification position - Adjustment locked6.9 Checking scale settings with respect to the calibration of weighing scalesFor the adjustment function, the balance must be switched over to the service mode.To achieve this, set the adjustment switch to adjustment position (See chpt. 6.8).In the service mode all parameters of the balance can be modified. The service pa-rameters may not be modified, as this could damage the balance settings.6.9.1 Menu overview in service mode (adjustment switch in adjustment posi-tion)This overview exists merely for the examination of the set parameters bythe authorized Verification Standards Department.Changes may only be made to the parameters of the automatic switch-off function“A.OFF“ and the acoustic signal ”BURR“.Navigation in the menu: With the weighing scales switched on, press the [→0←]-key for c. 3 sec until“SETUP“, followed by ”UNIT“ appears. Press the [ TARE ]-key repeatedly until the desired function appears. Confirm the selected function by pressing the [HOLD]-key. The first parameterappears. To select the desired parameter, use the [HOLD]-key and confirm bypressing the [TARE]-key.1 2 3