6.3 Mains connectionPower is supplied via the external mains adapter. The stated voltage value must bethe same as the local voltage.Only use original KERN mains adapters. Using other makes requires consent byKERN.6.4 Battery operationOn the rear side of the display unit remove the battery cover and connect 6 x 1.5Vmignon cells. Reinsert the battery cover.In order to save the battery, the balance switches automatically off after 3 minuteswithout weighing. Further shutdown times can be set in the Menu (Function"A.OFF"), see Section 9.If the batteries are exhausted, the battery symbol will appear (see chap. 2.1). Switch-off balance and replace batteries at once.If the balance is not used for a longer time, take out the batteries and store themseparately. Leaking battery liquid could damage the balance.6.5 Initial CommissioningIn order to obtain exact results with the electronic balances, your balance must havereached the operating temperature (see warming up time chap. 1). During thiswarming up time the balance must be connected to the power supply (mains,accumulator or battery).The accuracy of the balance depends on the local acceleration of gravity.Strictly observe hints in chapter Adjustment.EOB/EOE_L/EOE_XL/EOS-BA-e-1532 17