28 EWJ-BA-e-14118.3 OverviewFactory settings are marked by *.Menu block Menu item Available settings / explanationF1 UntWeighing Units Not documentedF2 blDisplaybackgroundilluminationEL AU* Automatic background illumination on whenweighing pate is loaded or key pressed.EL on Background lighting of display is switched onpermanentlyEO oFF Display background illumination offF3 ComInterfaceparameterS 232S USbUse to select interface: RS232 or USBP Prt* • Output of a stable weighing value after pressingof(menu setting „F4 ACC oFF“)• Manual totalizing (menu setting „F4 ACC on“),see chap. 7.9.Press and the weighing value will be addedto the summation memory and issued.P Cont Continuous data outputsee chap. 9.3P AUto For automatic add-up see chap. 7.10.This function is used to issue and add individualweighing values automatically to the summationmemory on unloading of weighing scale.wirel Not documentedP ASK Remote control instructionsCommand FunctionS Stable weighing value for the weight is sentvia the interfaceW Weighing value for the weight (stable orunstable) is sent via the interfaceT No data are sent, the balance carries outthe tare function.Z No data are sent, the zero-display appears.P Quantity will be sent via the interfaceAcknowledge selection by .b600b 9600*Baud rate600, 1200, 2400, 4800, 9600 selectable.Acknowledge selection by .