FCE-SH-e-0910 98 To Replace Load Cell Assembly1. Disassemble top housing of scale.2. Disconnect loadcell wires from main PCB. Remove the screws fixing the bottom plateof the load cell assembly. Replace the load cell with a new one.Connect load cell wires and fix the screws.3. Put 110% full capacity loading onto weight stand of scale. Adjust the overload screwson the weight stand until they just not touch the bottom plate.4. Assemble the top housing.5. Perform internal calibration as described in section 3.6. Check accuracy of scale at different weight.FCE 3K1 FCE 6K2 FCE 15K5weight tol. weight tol. weight tol.1000g 2g 2000g 4g 5000g 10g2000g 2g 4000g 4g 10000g 10g3000g 2g 6000g 4g 15000g 10gFCE 30K10 FCE 30K10L FCE 60K20weight tol. weight tol. weight tol.10.00kg 0.02kg 10.00kg 0.02kg 20.00kg 0.04kg20.00kg 0.02kg 20.00kg 0.02kg 40.00kg 0.04kg30.00kg 0.02kg 30.00kg 0.02kg 60.00kg 0.04kg7. Check other functions, such as Tare, Off and Auto-Off.