8 HFM-SH-e-11103.3 Setup of capacity (maximum load)1) Switch off the balance.2) Open the top cover of housing.3) Set jumper on pin J2.4) Switch on the balance and during the self test press the ZERO and thebutton once at the same time.„P3 - - - “ is displayed.5) Enter password „999“ and confirm with ZERO button.Pressing the button increases the numerical value of the blinkingdigit by one.Pressing the HOLD button moves the digit.6) The current capacity is displayed, e.g. 1000Press button to change the capacity and confirm with ZERObutton.7) Now balance will return to normal weighing mode.8) Switch off the balance and remove the jumper J2.