MCN-M-SH-e-2220 215.3 Adjustment procedure1. Turn the scale on and allow the scale to acclimatize and stabilize for sufficienttime.2. Press the verification button to enter Service Menu (may need to break theseal). Display shows X10.3. Press or to scroll to Adjust4. Press ➔, display show cal.5. Press ➔, display show calext.6. Press ➔, display show the first selectable calibration weight.7. Press or to scroll to other selectable calibration weights.8. Press ➔, display show zero, then ptld, then display calibration weight.9. Place the calibration weight onto the scale gently.10. Press ➔, display show wait, then remvld. Remove the calibrationweight. Scale will return to the weigh mode and display weight.Note:• In step 10, in case Wrong is displayed, this indicates a calibration error.Turn the scale off and then on and repeat the procedures.• Press to scroll. Press to esc/exit. Press ➔ to enter/confirm.• In case the scale will be used at different location, set the gravity ofadjustment location and gravity of use location. Follow the paths below in theService Menu:Service Menu ➔ Adjust➔ CAL➔ GRAADJService Menu ➔ Adjust➔ CAL➔ GRAUse