40 PFB-BA-e-174114 Instant helpIn case of an error in the program process, briefly turn off the balance and disconnectfrom power supply. The weighing process must then be restarted from the beginning.Help:Fault Possible causeThe displayed weight doesnot glow. • The balance is not switched on.• The mains supply connection has been interrupted(mains cable not plugged in/faulty).• Power supply interrupted.The displayed weight ispermanently changing • Draught/air movement• Table/floor vibrations• Weighing pan has contact with other objects.• Electromagnetic fields / static charging (choosedifferent location/switch off interfering device ifpossible)The weighing result isobviously incorrect • The display of the balance is not at zero• Adjustment is no longer correct.• The balance is on an uneven surface.• Great fluctuations in temperature.• Warm-up time was ignored.• Electromagnetic fields / static charging (choosedifferent location/switch off interfering device ifpossible)Should other error messages occur, switch balance off and then on again. If the errormessage remains inform manufacturer.