RFE-BA-e-1811 177.8 VerificationUUGeneral hints:According to EU directive 2014/31/EU balances must be officially verified if they areused as follows (legally controlled area):a) For commercial transactions if the price of goods is determined by weighing.b) For the production of medicines in pharmacies as well as for analyses in themedical and pharmaceutical laboratory.c) For official purpose.d) For manufacturing final packages.In cases of doubt, please contact your local trade in standard.After verification the balance is sealed at the indicated positions.Verification of the balance is invalid without the "seal".Verification notesAn EU type approval exists for balances described in their technical data asverifiable. If a balance is used where obligation to verify exists as described above, itmust be verified and re-verified at regular intervals.Re-verification of a balance is carried out according to the respective nationalregulations. The validity for verification of balances in Germany is e.g. 2 years.The legal regulation of the country where the balance is used must be observed!Balances with obligation to verify must be taken out of operation if:• The weighing result of the balance is outside the error limit. Therefore, inregular intervals load balance with known test weight (ca. 1/3 of the max. load)and compare with displayed value.• The reverification deadline has been exceeded.