RPB-BA-defsi-0510447.6 AdjustmentAs the acceleration value due to gravity is not the same at every location on earth,each balance must be coordinated - in compliance with the underlying physicalweighing principle - to the existing acceleration due to gravity at its place of location(only if the balance has not already been adjusted to the location in the factory). Thisadjustment process must be carried out for the first commissioning, after eachchange of location as well as in case of fluctuating environment temperature. To re-ceive accurate measuring values it is also recommended to adjust the balance peri-odically in weighing operation.Procedure when adjusting:For verified balances adjustment by means of a jumper is locked. In order to carryout adjustment the two contacts of the conductor plate must be shorted with a jumper(see chapt. 7.7.1).Observe stable environmental conditions. A warming up time of ca. 10 minutes isrequired for stabilization. Ensure that there are no objects on the weighing plate.Operation DisplaySwitch on balance and during self-checkpress the TARE keyPNUse the number keys to enter password„0000“:confirmarPN----ØCAL F1Ensure that there are no objects on theweighing plate.VnLQAD