MLB_C-BA-e-1431 6719 Instant helpFault Possible causeDisplay is not lit up. • The display unit is not switched on• The mains supply connection has beeninterrupted (mains cable not plugged in/faulty)• Power supply interrupted• Fuse has blownThe display does not changewhen a sample is being loaded • Sample dish / dish holder is fitted incorrectly.The weight display changesconstantly / the stability displaydoes not appear.• Sample dish has contact with wind protectiondevice or heated cover• Draught/air movement• Table/floor vibrations• Electromagnetic fields / static charging (choosedifferent location/switch off interfering device ifpossible)Incorrect measuring result • Check adjustment• No resetting to zero before loading the sampleMeasurement is taking toolong • Incorrect shut-down criteria setMeasurement isnot reproducible• Sample is not homogenous• Drying time is too short• Drying temperature too high (e.g. oxidationsample material, boiling point of sampleexceeded)• Temperature sensor soiled or defectiveDrying does not start • Heated cover open• The mains supply connection has beeninterrupted (mains cable not plugged in/faulty)