Care instructions ESTETICA E302 Reprocessing methods according to DIN EN ISO 17664 | 2.2 General instructions for the reprocessing of KaVotreatment centresManual drying ▶ Supply dry, clean and uncontaminated compressed air accordingto ISO 7494-2 only.▶ Blow off the outside and inside with compressed air until waterdrops are no longer visible.Automatic drying Usually, automatic drying is part of the cleaning/disinfection cycle ofthe washer disinfector and should not be carried out above 95 °C.Servicing ▶ Discard damaged products.Control and functional test ▶ Perform a visual inspection to check for damage and wear andtear.▶ Test products for function regularly.Packaging ▶ Use standardised packaging material. Sealing should not be un-der tension.▶ Use sterile goods packaging in order to ensure sterility.Sterilisation All sterilisable products are temperature-resistant up to max. 138°C.KaVo recommends sterilisation in a steam steriliser (autoclave) ac-cording to EN 13060 / ISO 17665-1, e.g. STERIclave B 2200 / 2200P (KaVo) or Citomat K-Series (Getinge).▶ Sterilise with a triple fractionated vacuum (B class steriliser);hold for 4 minutes; 134°C ± 1 °C; overpressure: 2.13 bar.Storage ▶ Reprocessed products should be stored protected from dust withminimum exposure to germs in a dry, dark and cool place.▶ Comply with the expiry date of the sterilised items.2.2.1 Overview ESTETICA E30 reprocessingReprocessing items/medicaldevicesCleaning Disinfection Sterilisa-tionmanual automatic manual automatic automaticSurfacesDeviceXXCushionXXLightXX See also:2 IfU KaVoLUX 540LED / EDI / MAIALights, handlesXXXXXDentist elementHandle TM table / handle cup S tableXXCannula triple-function handpieceXXXHolder pad/anti-slip matXXXXXHandpiece holderXXInstrument arm S tableXXAssistant elementConnection parts of the suction hoseXXXXPatient elementTumbler holderXXTumbler fillerXXPorcelain spittoon bowlXXXX7 / 40