Instructions for use Everest scan pro, Everest scan eco, Everest scan Base Camp | Typ 4101, 41025 Troubleshooting5 TroubleshootingMalfunction Cause RemedyEquipment is out of order No mains supply ▶ Switch on at the mains.▶ Provide power supply or checkthe plug-in connections.▶ Switch on main switch.▶ Check the main fuse in themains socket outlet and replaceif necessary.Projector lamp fails to illuminate du‐ring scanning.Faulty projector lamp ▶ Have a service technician re‐place the lamp.Measurement cannot be made. Initialisation failureFaulty rotate/tilt mechanismNo more space available on harddriveInternal temperature too highSevere vibrationsMisaligned camera or projector len‐ses▶ Restart measurement.▶ Check the rotate/tilt mechanismand contact a service technici‐an.▶ Archive the data by saving to CDor equivalent data carrier or de‐lete from hard drive.▶ Reduce ambient temperature.▶ Test ventilator and repair if ne‐cessary.▶ Do not jolt the equipment.▶ Readjust KaVo Everest Scan.▶ Contact a service technician.Surface scanned in poorly (grainyimage, parts missing)The surface reflects too much or istoo transparentThe surroundings in the measuringdevice are too brightThe object is outside of the measu‐ring range▶ Use suitable plaster or stumpvarnish and modelling wax.▶ Keep the door closed whilemeasuring.▶ Check position of object, correc‐ting if necessary.Soiled lens. ▶ Clean lens using lens brushPreparation limit unrecognisable ordifficult to recognisePreparation limit poorly shapedDie mounted too low▶ Prepare the tooth according tothe appropriate processing in‐structions (http://www.kavo-everest.com/De/zahntechniker/news/prospekte.asp?na‐vid=550067&lan=De#prospek‐te).▶ Mount die in a higher position.24/30