Instructions for use PHYSIO Evo/Evo F/One7 Reprocessing methods according to DIN EN ISO 17664 | 7.1 General care instructions7 Reprocessing methods according to DIN EN ISO 176647.1 General care instructionsCAUTIONProduct damage due to improper disinfection.Malfunctions.▶ Use disinfectant in accordance with the manufacturer's instructions!▶ Only disinfect by wiping!▶ Do not immerse product or product parts in liquids!Damage to the paint surfaces as well as plastics can arise from the wide variety ofmedicines and chemicals used in the dentist's office.Tests have shown that there is no surface protection available that works with all com‐mercially available agents.As damage to the surface by these agents is very much dependent on the exposuretime, it is essential to wipe the affected areas using a moist cloth.Any residue arising from disinfectants can be cleaned to a certain degree on paintedand plastic surfaces using neutral, nonabrasive rinses and cleansers.New painted surfaces that still cause water to bead can be cleaned with water andnonabrasive, mild cleansers.7.2 Cleaning and disinfection▶ Use disinfectants according to the manufacturer's instructions for use.▶ Observe the safety data sheet of the manufacturer of the relevant disinfectant.▶ Clean all surfaces with a soft cloth and an approved disinfectant.Approved disinfectants:▪ Mikrozid AF (liquid or cloths); manufacturer: Schülke & Mayr▪ FD 322; manufacturer: DÜRR Dental GmbH▪ Incidin Liquid; manufacturer: Ecolab▪ CaviCide; manufacturer: Metrex35 / 38