Instructions for use PROPHYflex 3 - 2018 for Sirona connection - 1.006.9920, - 1.006.99277 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing7.3.1 Manual disinfection of the the exteriorKaVo recommends the following products based on compatibility of the materials. Themicrobiological efficacy must be ensured by the disinfectant manufacturer.▪ Mikrozid AF made by Schülke & Mayr (liquid or cloths)▪ FD 322 made by Dürr▪ CaviCide made by MetrexConsumables required:Cloths for wiping the medical device.▶ Spray the disinfectant on a cloth, then wipe down the medical device and allowthe disinfectant to act according to the instructions of thedisinfectant manufacturer.▶ Follow the instructions for use of the disinfectant.7.3.2 Manual disinfection of the interiorThe efficacy of manual internal disinfection must be demonstrated by the manufac-turer of the disinfection agent. With KaVo products, use only disinfection agents thathave been released by KaVo with respect to the compatibility of materials (e.g. WL-cid / made by ALPRO).▶ Follow the instructions for use of the disinfectant.7.3.3 Mechanical disinfection of the exterior and interiorKaVo recommends washer disinfectors in compliance with EN ISO 15883-1 that areoperated with alkaline cleaning agents.The validations were conducted with the VARIO-TD program, the cleaning agentneodisher® MediClean and the neutralisation agent neodisher® Z.▶ For program settings as well as cleansers and disinfectants to be used, pleaserefer to the Instructions for Use of the thermodisinfector.▶ In order to prevent damage to the medical device due to residual fluid, make surethat the inside and outside of the device is dry after the end of the cycle. Removeany residual liquids from the interior and exterior of the medical device using com-pressed air.The drying procedure is normally part of the cleaning program of the thermodisin-fector.NotePlease comply with the instructions for use of the washer disinfector.7.4 Care products and systems - ServicingNOTICEImproper care.Malfunction or property damage.▶ Do not clean the PROPHYflex with oils or care spray.21 / 26