15CALIBRATIONIf you use the scale for the first time it is calibrated. Because of this it’s not necessary to calibrate it once again.You can re-calibrate the scale if it goes under environmental influences like dust or dirtHow to calibrate? (Warning: Information only for calibration service)**Calibration may only arrange from accredited calibration laboratory**1. The scale must be powered OFF. Press and hold [ZERO] first, then press [ON/OFF] while keeping the [ZERO]button depressed and then releases both keys. The display will show the A/D value (a series of random numbers).2. Wait for the stable indicator to be displayed. Press [UNIT] the display will show „0SAVE“ and then the displaywill return to the random A/D value. Place the full load kilograms weight(s) on platform. Wait for the stableindicator to display then press [UNIT], the display will show „0SAVE“. Calibration is complete; remove theweight(s). Turn the scale OFF (Press the ON/OFF key), then turn it back ON and check some weight readings.If calibration is still incorrect, repeat calibration but try it more slowly.If calibration is repeatedly incorrect then: 1) Replace the batteries, 2) Try a more stable surface or vibration andinterference free location, 3)Be sure there is no static electricity on the scale (use an anti-static spray to remove static electricity)FEATURES• Auto shut off• Display incl. wall holder• Hold-function• Weighing units: kg / IbSPECIFICATIONSModel number Capacity Division29924 150 kg 50 gAccuracy class IIICalibration value e = d = 50 gPlatform 330 x 320 mmNet/gross weight 4.1kg/4.5kgDisplay 6-digits, 15 mm high (LCD)Operation Temperature 15°C – 30°CPower source 6×AA Batteries or AC/DC AdapterCharger: 9~12V/100mA**WARNING**Cell-Phones, Cordless-Phones and any radio-frequency device can cause temporary interference and causethe scale to temporarily not work properly. While there is NO risk of permanent damage to the scale, interfe-rence can cause an incorrect calibration or incorrect weight readings.