Installation Note E8251-90382 5Add Option 1EH (Improved Low Band Harmonics from 10 MHz to 3.2 GHz) Kit Part Number:E8251-60382Safety ConsiderationsCheck Signal Generator FunctionalityUse the following procedure to confirm that the signal generator powers up and the internal checkidentifies no errors. The internal check evaluates the operation of the signal generator and returnsan error message if it detects a problem.1. Turn on the signal generator and let it warm up for at least five minutes.2. Run the signal generator self-test by pressing Utility > Instrument Info/Help Mode > Sel fTest > Run Complete Sel f Test. Upon completion, a summary of the self-test will bedisplayed. Use the E8257D/67D PSG Service Guide to troubleshoot any failures detected bythe test.3. Check to see if the ERR annunciator is on.— If the ERR annunciator is on, review the error messages in the error queue by pressingUtility > Error Info. The first page of error messages in the error queue appears in thedisplay text area. (Refer to the signal generator error messages document for informationabout each error message.)After resolving all problems causing errors, press Clear Error Queue(s).— If the ERR annunciator is off, the signal generator functionality check has passed.Before you disassemble the signal generator, turn the power switchoff and unplug the power cord. Failure to unplug the signalgenerator can result in personal injury.Electrostatic discharge (ESD) can damage or destroy electronic components. Allwork on electronic assemblies should be performed at a static-safe workstation.When the signal generator is first connected to ac line power, the error messageERROR 514, Reference Oven Cold occurs which causes both the OVEN COLDannunciator and the ERR annunciator to turn on.After approximately five minutes, the OVEN COLD annunciator automaticallyclears, but the ERR annunciator remains on until all errors are cleared from theerror queue.Some circuits may require up to 50 minutes to warm up before passing theself-test. If self-tests continue to fail after 50 minutes of warm up, troubleshootthe instrument.