Installation Note N9030-90090 5Option B2X, or Option BU8, 255 MHz Analysis BandwidthInitial Instrument Functionality Check1. Power on the instrument and allow the instrument to boot up, run the alignments and displaythe measurement screen. The instrument will probably display a spectrum analyzer screen andyou will see the instrument sweeping.2. There should be no alignment failures. If there are failures, investigate and fix the problembefore continuing.Installation ProcedureAnalyzer Information1. Connect a power cord to the analyzer and turn on the analyzer.2. After the analyzer has completed turning on, press System, Show, System.Make note of the following information from the Show System screen:Product Number _______________Serial Number _______________Instrument S/W Revision _______________3. Check for the presence of the options listed below in the Show System screen. Put a checkmark after each option listed below that appears in the show System menu.N9030B-526, or 513, or 508 _______________N9030B-MPB _______________N9030B-B85 or B1X _______________4. Refer to the data in step 2 above. Verify that the Product Number in step 2 is appropriate forthe upgrade being installed:If the Product Number in step 2 is not appropriate for the Option B2X/BU8 upgrade, do notproceed with the installation.5. Refer to the data in step 2 above. If the Instrument software is earlier than A.18.24, you mustupgrade to version A.18.24 or later. Keysight recommends that you update to the latestinstrument software version to ensure that you have the latest defect fixes. To check the latestinstrument software version, visit the following website:http://www.keysight.com/find/N9030B_software6. Refer to the data in step 3 above.Verify that N9030B-526, or 513, or 508 is checked (currently installed). This retrofit kit is forinstruments with frequency range ≤ 26.5 GHz. It will not work on 3.6 GHz, or 44 GHz or 50 GHzinstruments.Kit to be Installed Product Number (Step 2)N9030BU-B2X or BU8 N9030B