22 Keysight E1406A Command Module Service GuideVerification Tests Command Module ConfigurationSome Command Module DefinitionsIn this manual, the VXIbus “system” assumed is an external controller connectedto a command module via GPIB, with the command module configured as theresource manager and the slot 0 device. The following paragraphs summarizesome terms associated with the command module. See the C-Size VXIbusSystems Configuration Guide for details.Resource Manager/Slot 0 DeviceEvery VXIbus system must have a device to provide the system’s resourcemanager and slot 0 requirements. The resource manager operates ONLY atpower-on. Once the power-on sequence completes, the resource manager is nolonger used. The resource manager:– identifies all installed plug-in modules– sets commander/servant hierarchies– performs A24/A32 mapping– allocates interrupt lines– starts system operationDuring operation, the slot 0 function is used to:– identify module locations– manage the data flow across backplane buses– provide the (10 MHz) system clock.Logical Address/GPIB AddressFor VXIbus systems, the Logical Address (set with the Logical Address Switch) isused to:– determine device registers base address– set a device as the system resource manager– establish servant areas– create instruments– derive secondary GPIB addressesThe GPIB (IEEE-488) Address is used to address the device and consists of thefollowing three parts:– Interface Select Code (ISC) (typically 7)In VXIbus systems using an external controller, the KeysightE1406A Command Module should be configured as resourcemanager and slot 0 device.