7 8Should you wish to change the temperature scale from centigrade to Fahrenheit enter themenu as before and use the arrow icons/buttons until you get to a screen displaying C F.Touch the tick icon/button to enter this function as before.Centigrade/FahrenheitH/LPWhen you access the menu the first menu option is to view the set point, see above.The set point is one of the parameters that you can not change but you can view it. YourLabcold IntelliCold® pharmacy refrigerator leaves the factory with a set point of 5oC as advisedby the Green Book chapter 3 page 27 and to prevent this being altered accidently, the setpoint can only be changed by a Labcold engineer. If you need the set point changed pleasecontact the Labcold service team on 01256 705 580 or service@labcold.comTo view the set point, touch the tick icon/button when the SEt screen is displayed and thecontroller will display the set point temperature.Viewing the Set PointH/LP°CH/LPTo return to the normal display, touch the muteicon/button or wait 30 seconds.Viewing Software VersionThis is another function that you cannot change, but you may be asked about the softwareversion if you need to contact us about an issue.If the controller has reverted to the normal display, hold the arrow icons/buttons as above,if you are already in the menu, touch the up and down arrows as required until you get to ascreen withSFt on it. Touch the tick icon/button as before and it will display a number. Again touch themute icon/button or wait 30 seconds for the display to revert to normal.H/LPH/LPH/LPThe C will flash, touch one of the arrows until the display shows F. Hold the tick icon/buttonto confirm your selection and the screen will change from F to the the normal temperaturescreen. The temperature will now display in Fahrenheit.The temperature data for the Micro SD card will always be saved using the centigrade scale.°°t 1 FH/LPAlarm Time OutThe alarm time out function allows you to change the time interval between the alarm beingmuted and the alarm sounding again.Access the menu by holding the arrow icons/buttons simultaneously and use either one ofthem to scroll through options until you get to a screen that shows Ato.Touch the tick icon/button to enter the function. The display will now flash with the currentnumber of minutes there are between the alarm being muted and sounding again. The fac-tory setting is 15 minutes as shown below.H/LPUse the arrow icons/button to adjust the interval in 5 minute units up to a maximum of a 30minute time out. When you have set it to the length of time you require, hold the tick icon/button and the screen will revert to Ato. You can then either leave it alone or hold the muteicon/button and the controller will then revert to the normal temperature display.H/LP