Chapter 4: Using Your CentriVap Mobile SystemProduct Service 1-800-522-7658 19If during a run the “STOP” button is pressed, the timers remember the time atwhich stop occurred. If “RUN” is then pressed, the timers continue to countdown from the time at which they were stopped. If you are running program 1,program 2 or program 3, pressing “PROG 1,” “PROG 2” or “PROG 3” resets thetimers to the original set point time and the CentriVap starts a new run.To reset the timers to the original set point time when running programs otherthan program 1, program 2 or program 3, press the start button and hold it for fiveseconds. The display will indicate that the timer is reset.Operating the CentriVapCAUTION: To avoid personnel injury; Do not operate the CentriVap if thelid is scratched or nicked, or shows signs of damage. A damaged lid could failunder vacuum.While the CentriVap Concentrator is operating, do not lean on the lid, do notstand near it longer than necessary and do not place hazardous materials within 12inches.1. Turn “ON” the CentriVap power switch.2. Press the Cold Trap “ON” switch. The “Y” in the display under the cold trapwill flash. When the trap reaches operating temperature the “Y” willilluminate steady.3. Select a program or set the set point parameters.4. Preheat the chamber if desired.5. Turn on Heat Boost if the CentriVap is so equipped and additional heat isdesired.6. Place samples in vials. Normally the vials should be no more than half full.Place vials in a rotor.7. Load the rotor with samples into the chamber.8. Close the lid.