Command DescriptionIN_SP_05 Query of outflow temperature limit LoIN_PAR_00 Query of the control parameter XpIN_PAR_01 Query of the control parameter Tn (181 = OFF)IN_PAR_02 Query of the control parameter TvIN_PAR_03 Query of the control parameter TdIN_MODE_00 Keyboard master: 0 = free / 1 = blockedIN_MODE_02 Standby: 0 = device ON / 1 = device OFFTYPE Query of the device type (answer = "LOOP")VERSION Query of the software versionSTATUS Query device status 0 = OK, -1 = faultSTAT Query of fault diagnosis response: XXXXXXX; X = 0 no fault, X = 1fault1st Character = error2nd Character = not assigned, always 03rd Character = not assigned, always 04th Character = not assigned, always 05th Character = sublevel6th Character = not assigned, always 07th Character = not assigned, always 0Take the following information into account: " " (blank space) can also be used for "_". If not otherwise specified in the command, the response willalways be given in fixed point format "XXX.XX" or for negativevalues "-XXX.XX" or "ERR_X".5.11.5 Error messagesThe following is a description of the error messages of the RS 232interface.Error DescriptionERR_2 Wrong entry (for example, buffer overflow)ERR_3 Wrong commandERR_5 Syntax error in valueOperationLOOP34