Troubleshooting TipsWhy is the wireless data logger not appearing in the software?If the RFPulse2000A doesn’t appear in the Connected Devices panel, or an error message isreceived while using the RFPulse2000A, try the following:• Check that the RFC1000 is properly connected. For more information, see Troubleshooting InterfaceCable problems (below).• Ensure that the battery is not discharged. For best voltage accuracy, use a voltage meter connected tothe battery of the device. If possible, try switching the battery with a new 9V lithium.• Ensure that no other MadgeTech software (such as MadgeTech 2, or MadgeNET) is running in thebackground.• Ensure that MadgeTech 4 is being used. MadgeTech 2 and MadgeNET are not the correct software forconnecting with the RFPulse2000A.• Ensure that the Connected Devices panel is large enough to display devices. This can be verified bypositioning the cursor on the edge of the Connected Devices panel until the resize cursor appears, thendragging the edge of the panel to resize it. The screen layout may also be reset in the options menu byselecting File, Options, and scrolling to the bottom.Troubleshooting Interface Cable problemsCheck that the software properly recognizes the connected RFC1000 wireless transceiver.If the wireless data logger is not appearing in the Connected Devices list, it may be that the RFC1000 isnot properly connected.1. In the software, click the File button, then click Options.2. In the Options window, click Communications.3. The Detected Interfaces box will list all of the available communication interfaces. If the RFC1000 islisted here, then the software has correctly recognized and is ready to use it.Check that Windows recognizes the connected RFC1000 wireless transceiver.If the software does not recognize the RFC1000, there may be a problem with Windows or the USB drivers.1. In Windows, click Start, right-click Computer and choose Properties or press Windows+Break as akeyboard shortcut.2. Click Device Manager in the left hand column.3. Double click Universal Serial Bus Controllers.4. Look for an entry for Data logger Interface.5. If the entry is present, and there are no warning messages or icons, then windows has correctlyrecognized the connected RFC1000.6. If the entry is not present, or has an exclamation point icon next to it, the USB drivers may need tobe installed. These are available on the software flash drive included with the RFC1000, and on theMadgeTech website.Ensure that the USB end of the RFC1000 is securely connected to the computer.1. Locate the USB-A plug of the RFC1000.2. If the interface cable is connected to the PC, unplug it. Wait ten seconds.3. Reconnect the cable to the PC.4. Check to make sure that the red LED is lit, indicating a successful connection.MadgeTech, Inc.6 Warner Road l Warner, NH 03278Phone 603.456.2011 l Fax 603.456.2012www.madgetech.com l info@madgetech.comREV 2 2013.12.20Distributor: LoggerShop Technologies | DorsetTel: +44 (0)845 520 0202 Fax: +44 (0)845 520 0203 | Email: sales@loggershop.co.uk | www.loggershop.co.uk