7-24 Remote control7.7 Sweep ModecommandssequenceexampleThis is the sequence used by the PMM Emission Suite when a scantable is run, inthe “conducted” range (Band B from 0.15 to 30 MHz).ExampleCommand sent Reply received Description#?IDN* Queries information about model, release and date of the firmwareIDN=7010/03-FW - 1.09 11/06/14 It is a 7010/03 with its FWIt is used to check congruencies, frequency limits, and all the relatedcharacteristics#?S/N* Queries the serial numberS/N=000WE20304 Serial number#?CRA* Queries if RMS-AVG optional detector is activeCRA=OK RMS-AVG active#SCFA -1* Correction factors, if present, deactivatedCFA=OK (OFF) Important. Could be previously stored and activated.#S3PRC* Sets the unit in the bands A and B for conducted3PR=OK ack#SSFD150000;30000000;5000;P;2;6;10;OFF;OFF;0*SweepBinary data reply Measurements data#?UHT* Queries the hold timeUHT=2.5ms Applied hold time