BEFORE ANY HANDLING:• Disconnect the AC power cord.• Check that all electrical components are connected to each other.• Clean electrical enclosures of the actuators and of the handsets, if they have been soiled by bodyfluids, especially urine.OBJECTIVE:• Repair the stand-up lift and prevent the transmission of germs from one patient to another.• Remove any organic soil by mechanical action (cleansing) or chemical action (disinfection).CLEANING PROCESS:• Clean the surfaces with a wet cloth by using an apropriate detergent.• A regular cleaning is recommanded.WARNING:• The detergents used must have a neutral pH.• Avoid the abrasive products and solvent, they could damage the device surface state.DAILY MAINTENANCE:• By means of a product applied on the surface in one operation.• After the patient’s discharge by respecting the following operation:- Cleaning is carried out by means of a cloth impregnated with a disinfectant solution surface.• Maintenance by specific service providers after removal of the stand-up lift from the facility:- Biological cleaning operation- Steam cleaning of various flat surfaces. Change surfaces regularly when washing to avoid waterload. Steam cleaning of inaccessible areas. For tubes, steam use with a microfiber mop. Do not directsteam directly on the electrical boxes.CAUTION:• Disinfection of actuators, electrical boxes and handsets with a microfiber cloth impregnated withdisinfectant.MONTHLY MAINTENANCE:• Check that the device does not present any visible damages.• Check that no part is missing.• Check the good functioning of the castors and that no matter prevents their bearing.• Check the good fonctioning of the commands as well as the connections of the remote controland the actuator on the battery.• Clean the sockets and the commands buttons with a dry cloth, if needed with a wet cloth.• Check the good state of the electric cables.Never clean the stand-up lift, especially the electrical system, with a high pressure cleaner, waterhose or similar because doing so could damage the surfaces and seals, and water could enter.Cleaning & Maintenance19User Manual - Stand-Up Lift: WAYUP 4 with electrical-opening legsManufactured by NAUSICAA Médical S.A.S. / Approved by Ghizlane Labrosse (Biomedical Engineer)