Model 6700 Family User’s Manual 13The Diagnostic (DIAG) mode menu allowstesting of specific areas of the scale’s functionand viewing current configuration settings. Thisincludes the following:DISPLAY (DISP) – Shows the version andrevision of the software, followed by a displaysegment test.RAM (RA) – Performs a nondestructive test ofRAM in the processor.ROM (RO) – Performs a checksum of all loca-tions of ROM in the processor.INPUT/OUTPUT (I/O) – Data is output by thescale and through the use of a loopback connec-tor. The data is immediately read back into thereceive channel and verified against what wassent. Requires a jumper (short) between transmitand receive data lines.DIVISION TEST, w/AZT (DIV-A) – Weight datais normalized to 100,000 counts of displayedresolution. AZT (Auto Zero Tracking) is enabled.DIVISION TEST, w/o AZT (DIV-N) – Weight datais normalized to 100,000 counts of displayedresolution. AZT is disabled.Viewable configuration settings include:LN FR Line frequencyFILT Filter settingPROT Serial protocolBAUD Baud rateCAP Scale capacityUNITS Unit of measureDiagnostics ModeDiagnostic (DIAG) ModeTip: Easy access tothe diagnostic mode isavailable directly fromthe front panel withoutopening the scale toaccess the Menumode switches. SeeSetup Review/Testsection.