Model 7600 Family PWC User’s Manual 91. Check container for any obvious evidence ofdamage.2. Remove contents of the shipping container.3. Inspect the scale for shipping damage.Immediately report any damage to theshipper.1. Mount the scale on a stable, level surfacefree from air currents and vibration. Be surethe scale platter does not touch any adjacentsurfaces.2. To install the scale surface flush with acountertop, use these dimensions to guideconstruction:Platform Minimum CutoutDimensions Dimensions7620 14" W 14.75" W12.5" D 13.25" D4.2" Min Ht.7680 18" W 18.75" W18" D 18.75" D4.6" Min Ht.3. Loosen the plastic collars (7620) or jam nut(7680) on the leveling feet. Level the scale byusing the level bubble under the scale platteras a guide. Be sure all four feet are in firmcontact with the counter, then tighten allcollars (nuts).Initial SetupUnpacking the ScaleInstalling the Scale