Model 7821 Bench Scale User’s Manual14The Configuration (CONF) Mode menu allowsconfiguration of the scale to the specific applica-tion need. The items you can configure are asfollows:Filtering (FILT) – Choose between FASTand SLO filtering. Slow should be chosen inareas susceptible to vibration. Choose FASTfiltering for more stable conditions.Baud (BAUD) – Choose a baud and parityfrom table.Protocol (PROT) – Choose serial communi-cation protocol:NCI - Ref. 8408-14788-01 (W, S, Z, H, U, MOnly)8213 - Ref. 8408-14788-033835 - Ref. 8408-14788-12Follow these steps to access and configure theitems in the CONF menu. Refer to Figure 2.1. From the normal weighing mode, moveSwitch 1 to Menu Mode or OPEN position,then press the ZERO key repeatedly until…CONF is displayed.2. Press the UNITS key…FILT is displayed. This stands for filter-ing.3. Press the UNITS key…The current setting, FAST or SLO, isdisplayed.10Configuration ModeStep-by-StepInstructions forCONF Mode