English-4Intended UseThe MD211G5 grayscale display is intended to be used for displaying and viewing of digital images for diagnosis, includingdigital mammography, by trained physicians.Caution: The MD211G5 grayscale display cannot be used for a life-support system.This device is designed as component of a fi nal system which is in compliance to IEC 60601-1-1 requirements.Contraindications: None known.Safety PrecautionsThis device is designed for exclusive interconnection with IEC 60950 certifi ed equipment outside of patient vicinity.• Equipment connected to digital interfaces must comply with the respective IEC standards (e.g. IEC 60950 for data processingequipment and IEC 60601-1 for medical equipment).• This device complies with IEC 60601-1-2. To minimize the interference from other equipment, a minimum 0.5 m distance shallbe kept form other potential electromagnetic sources, such as a Cell Phone.• To reduce the risk of electric shock, make sure that the power cord is unplugged from wall socket. To fully disengage thepower to the unit, please disconnect the power cord from the AC outlet. Do not open the back cover. No user serviceable partsinside. Refer servicing to qualifi ed service personnel. The AC outlet shall be readily available and accessible.EMC PrecautionThis product requires special EMC precautions and needs to be installed and used according to the following information.Use the specifi ed cable. Using other cables could increase emission or decrease immunity.Do not place any RF communication equipment in the vicinity of this product. It could affect the performance of the product.Any additional equipment connected to this product needs to comply with the requirements of IEC 60601-1-2.North America CustomersGrounding reliability can only be achieved when the equipment is connected to an equivalent receptacle, marked “Hospital Only”or “Hospital Grade”. That signal input port need to be connected properly and that any unused signal input port shall not beaccessible for the patient to contact in the patient area after the LCD is integrated into a medical system.Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner.European CustomersUnpacking, Installation and calibration of this display should only be done by authorized and trained personnel. Any installationdone by a non-authorized person is done under his own risk and we accept no responsibility in any device malfunctioning. Tosecure full functionality of this display, only NEC approved display controller models for connection with the fi nal system must beused.Medical ImagingMD211G5 is designed for 2048 x 2560 medical imaging.