Appendix A Specifications© National Instruments Corporation A-3 NI PXI-8184/8185 User ManualShock and VibrationOperational shock .................................. 30 g peak, half-sine, 11 ms pulse(Tested in accordance withIEC-60068-2-27. Test profiledeveloped in accordance withMIL-PRF-28800F.)Random vibrationOperating ........................................ 5 to 500 Hz, 0.3 grms(with solid-state hard drive)Nonoperating .................................. 5 to 500 Hz, 2.4 grms(Tested in accordance with IEC-60068-2-64. Nonoperating testprofile exceeds the requirements of MIL-PRF-28800F, Class B.)Note For EMC compliance, operate this device with shielded cabling. In addition, allcovers and filler panels must be installed. Refer to the Declaration of Conformity (DoC)for this product for any additional regulatory compliance information. To obtain the DoCfor this product, visit ni.com/certification, search by model number or product line,and click the appropriate link in the Certification column.SafetyThis product is designed to meet the requirements of the followingstandards of safety for electrical equipment for measurement, control,and laboratory use:• IEC 61010-1, EN 61010-1• UL 61010-1• CAN/CSA C22.2 No. 61010-1Note For UL and other safety certifications, refer to the product label or visitni.com/certification, search by model number or product line, and click theappropriate link in the Certification column.