Related DocumentationFor additional information, refer to the following documents as you perform the calibrationprocedure:• PXIe-5413 Getting Started Guide• PXIe-5413 Specifications• NI Signal Generators Help• LabVIEW HelpVisit ni.com/manuals for the latest versions of these documents.Test EquipmentRefer to the following table for a list of necessary test equipment, minimum requirements forthe test equipment, and model recommendations for calibration of the PXIe-5413.If you do not have the recommended equipment, select a substitute calibration standard usingthe minimum requirements specifications listed in the table.Table 1. PXIe-5413 Test EquipmentEquipment RecommendedModelWhere Used Minimum RequirementsOscilloscope PXIe-5160 orPXIe-5171Verifications:• TimebaseaccuracyAdjustments:• TimebaseFrequency resolution:0.1 HzThe ability to measure sinewaves ≥10 MHz withamplitudes of 2 Vpk-pk to5 Vpk-pk with 50 Ω inputimpedance.SMA (m)-to-SMA (m) cable— Verifications:• TimebaseaccuracyAdjustments:• TimebaseFrequency range: DC to100 MHzImpedance: 50 ΩLength: <1 m2 | ni.com | PXIe-5413 Calibration Procedure