A-1APPENDIX A SERVICE MODEA.1 SERVICE MODEThis section of the manual contains service mode commands.A.1.1 Service Mode Commands:The Service Commands are only enabled after the scale receives a SV command through the RS-232 interface. This command sets the scale into the Monitor mode. The Monitor mode is exited whenthe scale is turned off with the ON/ZEROOff button or a power off/on cycle, or a command is receivedwhich resets the scale. (The Monitor mode is exited regardless of the state of battery power.) Table A-1 lists the available service commands.TABLE A-1. SERVICE MODE COMMANDS.SV Enters the Monitor ModeRS Reset the scale, (similar to power off and on) and exits the Monitor ModeEPX* Indicator is set to 10 times expanded display mode. X=1 ON, X=0 OFFDF* Resets all scale set ups to default values for a CKW55.BD Battery voltage, temperature, and charge state are printed. (x.xxV yy°C zzzCC w,where zzz= the recorded number of full charge cycles, w=the present chargestate 0 to 4, 0=off, 1=precharge 2=fast, 3=trickle, 4=maintenance.) Indicates sending the ASCI code for the ESC key (27).*The Service Commands marked with a *, will not be operational if the LFT menu selection has been setto On, or if the LFT “LOCK” switch has been set to “LFT” (“LOCK”) position.NOTE: All commands sent to the Indicator must be terminated with a carriage return (CR) or carriagereturn-line feed (CRLF).