CHAPTER 4 - TESTINGOhaus Corporation www.ohaus.com 46 Explorer® Series Service Manual4.2 OPERATIONAL TEST1. Connect a functioning Power Adapter to the balance.2. Plug the Power Cord into a suitable power source.3. Verify that the balance start up sequence occurs properly. Note any error codes.4.3 LOAD CELL TEST USING RAMPTo test the Load Cell using RAMP, see Appendix B.4.4 PERFORMANCE TESTSAccurate performance of the Explorer ® balances is determined by a series of four performancetests. The displayed readings are compared with the tolerances listed in Tables 1-1. Tolerancevalues are expressed in counts. A one-count difference is shown in the last digit on the balancedisplay.The following performance tests are used to evaluate balance operation before and after repairs.The balance must meet the requirements specified in each test as well as the specifications listed inTable 1-1. Before proceeding with the following tests, the balance should be warmed up (60minutes) and calibrated. (See Appendix A)4.4.1 Precision TestThe Precision Test is a quick test that measures the deviation of a limited number of weightreadings. If the balance passes the precision test than the following tests should be performed.Note: This is a reference test for approved models (OIML and NTEP). It is not a required test but itcan be used to determine if the balance is working properly.1. The reading on the display should be 0g.2. Select a mass weighing near the maximum capacity of the balance, and place it on the centerof the Pan. Observe and record the reading.3. Remove the mass. The reading should return to 0g ± 1d.4. Repeat this test three times. The readings should be within ± 1d. If so, the balance passes thePrecision Test.If the deviation for any set of readings (using the same mass placed on the center of the Pan) isgreater than 1d, the balance does not meet the precision specification. Inspect and correct thefollowing areas:- Check for mechanical obstructions. Any foreign object touching any part of the movingassemblies will cause a balance to fail the Precision Test. Inspect and correct as necessary.- If the balance does not meet specifications, move it to a suitable location, ensure that it islevel, and try again. If it still does not meet specifications, perform a service calibration, and tryagain. (See Appendix B for Service Calibration)- If the balance does not pass this test, the Load Cell may need to be replaced.