Ohaus Corporation www.ohaus.com 57 Explorer 5 Semi-Micro BalancesTABLE C-1 INTERFACE COMMAND LIST (CONTINUED)Command Functionx% Set Percent application reference weight (x) in grams. (must have reference weight stored)P% Print Percent application reference weight.xAW Set Dynamic Weigh Level to x. (x = 1 - 99 seconds)xAW Set Dynamic Weigh Mode. x = A (Automatic), S (Semi-Automatic), M (Manual)PAW Print Dynamic Weigh Level.BAW Start Dynamic Weigh cycle. (Manual Mode)CW Clear locked weight (weight < threshold) in Dynamic Weigh, Display Hold & Totalize.xCO Set Checkweighing Over Limit in grams x.xCU Set Checkweighing Under Limit in grams x.PCO Print Checkweighing Over Limit.PCU Print Checkweighing Under Limit.xCM Set Checkweigh mode (1=over/under, 2=target/% tolerance, 3=Target/weight tolerancexCT% Set Checkweighing target in grams x for percent tolerance mode.PCT% Print Checkweighing Target for percent tolerance mode.xCTW Set Checkweighing target in grams x for weight tolerance mode.PCTW Print Checkweighing Target for weight tolerance mode.xC% Set Checkweighing % tolerance x. Attention: when x is a positive value, it is used to set the+tolerance value; vice versa.PC% Print Checkweighing % tolerance.xCW Set Checkweighing weight tolerance x. Attention: when x is a positive value, it is used to set the+tolerance value; vice versa.PCW Print Checkweighing weight tolerance.xDH Set Display Hold mode (Peak Hold). x = A (Automatic), S (Semi-Automatic), M (Manual)xD Set 1 second print delay (set x = 0 for OFF, or x = 1 for ON).xFL Set filter level to x (1 = low, 2 = med, 3 = high).xAL Set Auto-zero to x (x = 1 for 0d, x = 2 for 0.5d, x = 3 for 1d, x = 4 for 3d).Esc R Resets all Balance menus to factory defaults. Attention: The binary code of this commands is “1B20 52 0D 0A” or “1B 52 0D 0A”.PID Print current user Name.xID Program user Name. Attention: only allowed numeric input.xTL Set Totalize Mode. x = A (Automatic), M (Manual).PTIME Print current time.PDATE Print current date.xTIME Set Time, x format: hh:mm:ss.xDATE Set Date, x format: mm/dd/yyyy.CA Continuous weight, same as CP.SA Stable load, same as SLP.xA Interval Print x = interval in sec (1-3600) 0 = off, same as xP.0A Set AutoPrint off, same as 0P.SC Begin Span Cal, same as C.xAM Set Animal Mode to Auto, Semi-Auto, Manual. Same as xAW(A/S/M)? Prints current mode, same as PM.xS 0 = print unstable data, same as IP; 1 = print stable only1), same as SP.xRL 0 = disable response; 1 = enable response. Attention: this command only controls the “OK!”response.