FD Series EN-203.8 Legal for Trade SettingsThe FD Series is designed to comply with OIML, EEC, NTEP and Measurement Canada weights andmeasures regulations. These approvals may be pending; contact Ohaus for further details aboutavailability.Note:Consult your local Weights and measures office for their regulations before placing the scale into aLegal for Trade application.When the LFT menu item setting is On, the following conditions apply:• The Calibration (CAL) sub-menu is hidden.• The Calibration Lock (LCAL) menu item is hidden.• The Auto Zero Tracking (AZt) menu item is set to 0.5 d.• lb:oz unit is disabled.When the weights and measures official has approved the scale for use, it should be sealed using oneof the methods shown in Section 3.9.3.9 Sealing the ScaleWhen used in conjunction with the Lockout menu and the security switch, the scale may besealed to prevent or detect unauthorized changes to scale settings. For Legal for Tradeapplications, set the LFT menu item to ON, set the security switch to the On position, then seal thescale to prevent access to the metrological parameters. Seal the scale according to the localweights and measures regulations.To regain access to the scale settings, break the seal and set the security switch to the OFF position.On OffFigure 3-5. Sealing methods.Security SwitchFigure 3-4. Security Switch Settings.Wire Seal Paper SealWire SealRemove the spacers under the cross drilledscrews. Feed the wire through the crossdrilled holes.Paper SealPlace the paper seals over the screws andsealing spacers.