FrontierTM Centrifuge FC5706 EN-114.2 Access to mode "Operating Data"When using the centrifuge, the following parameters can be set:-Number of starts-Operating hours of centrifuge-Operating hours of motor-Software-version-Error list-Function of the imbalance switch-Operation of keyboard-Display testsWhile the centrifuge is turned off, press simultaneously the keys "Time"(6) and "Door Open" (7) and turn on themain switch of the centrifuge. Now release both keys and as a result a display test is executed for approx. 5seconds. All indicators will appear at the same time (See figure 26).Figure. 26ATTENTION:After the settings have been stored, change back to normal mode again by switching off the centrifugefor a short while!4.3 Call up operating dataATTENTION:This should only be performed by advance user or service engineer.In the mode "Basic Adjustments" you can call up the operating data of the centrifuge. Please proceedas described under point 4.1.2 to enter this program mode. Press the key "Accel/Decel" (5). In the display"Acc/Dec" (A-2) the word "Service" flashes.With the adjusting knob (1) various information can be called up:A= previous starts of the centrifugeH= previous operating hoursS= software versionr= converter softwareE= list of previous error messagesh= running time of the motorThe list of the last 99 error messages can accessed by pressing the key "RPM | RCF" (4). Scroll through the list byusing the adjusting knob (1). The respective error codes appear in the display "RPM | RCF" (A-1). Please refer to"Table 5: error messages" (see APPENDIX).67