Quick Start Guide NavigatorTMOhaus Corporation www.ohaus.com 3 NavigatorTM4. INITIAL CALIBRATIONThis quick start guide explains how to use the basic functions of the instrument. Forinformation about advanced features, refer to the instruction manual.When the balance is first installed it should be calibrated to ensure accurate results.Press and hold Menu until [mMeNU] (Menu) is displayed. When the button is released,the display will display [.C.A.L.]. Press "Yes" to accept, [SpaN] will then be shown. Press"Yes" again to begin the span calibration.[--C--] blinks while zero reading is stored. Next, the display shows the calibrationweight value. Place the specified calibration mass on the pan. [--C--] blinks while thereading is stored. The balance returns to the previous application mode and is ready foruse.5. OPERATION5.1. ControlsButton FunctionsZeroYesShort Press (when on): Sets display to zero(when off): Turns balance onLong Press (when on): Turns the balance offShort Press (in Menu): Selects/accepts displayed settingPrintUnitsNoShort Press: See Interface Manual for operation description.Long Press: Toggles through active unitsShort Press (in Menu): Toggles through available settingsFunctionModeBackShort Press: Selects function settingLong Press: Selects active ModeShort Press (in Menu): returns to previous settingsTareMenu-CalExitShort Press: Enter / clear a Tare valueLong Press: Enters User MenuShort Press (in Menu): Quickly exit User Menu5.2. Menu StructureSub-Menus: .C.a.l. .S.e.t.u.p.* .M.o.d.e. .U.n.i.t.* .E.n.d.Menu Items:SpanLinEndA-offDisplyBrightIREndCountPercntCheckEndct, g, kg,grn, lb, oz,lb:oz, ozt,N, dwt,thk, tsg, ttw,tola, ticalEnd* Note: Available settings vary by models and regions.