Ohaus Corporation www.ohaus.com 14 Ranger 7000® Series Service Manual3 MAINTENANCE / REPAIR PROCEDURES3.1 PREVENTIVE MAINTENANCEOhaus scales are precision instruments and should be carefully handled, stored in a clean, dry,dust-free area, and cleaned periodically. Follow these precautionary steps:• When a scale has had chemicals or liquids spilled on it, all exterior surfaces should becleaned as soon as possible with warm water on a damp cloth.• Do not leave a mass on the scale when the scale is not in use.• Allow time for the scale to stabilize after moving it from an area which is at a differenttemperature than the area where it is to be operated. Allow one hour for each 5°F (2.7°C)temperature change before using the scale. After temperature stabilization, allow anadditional 60 minutes after turning the scale on, for the scale electronics to stabilize.Preventive Maintenance ChecklistThe scale should be inspected and checked regularly, as follows:1. Remove the Pan and Sub Pan to inspect and clean the area beneath the Pan.2. Clean the outside of the scale using a damp cloth with warm water.3. Check the Power Cord for broken or damaged insulation.4. Make a visual inspection for faulty connectors, wiring, and loose hardware.CAUTIONDO NOT USE CHEMICAL CLEANERS OR SOLVENTS OF ANY TYPE.SOME CLEANERS ARE ABRASIVE AND MAY AFFECT THE SCALE’S FINISH.