Additional Instructions Scout SJX BalancesOhaus Corporation www.ohaus.com i Scout SJX BalancesUNPACKINGCarefully remove your Scout balance and each of its components from thepackage. Store the packaging material to ensure safe storage and future transport.Included Components: Balance Pan User Guide &Instructions Calibration Mass (only for SJX/E models with 64gcapacity) Warranty Card Scoop (only for models with 64g capacity) AC Adapter & PlugCALIBRATIONWhen the balance is operated for the first time, a calibration is recommended toensure accurate weighing results. External Span CalibrationWith balance turned on:Step-1 Press and hold Menu until [mMeNU] (Menu) is displayed.Step-2 Release the button, the display will show [C.A.L].Step-3 Press Yes to accept. [SpaN] will then be shown.Step-4 Press Yes to begin the span calibration. Zero reading is storedautomatically.Step-5 Place the specified span calibration mass on the pan. Then, press Yesto accept the weight or No to select an alternate weight.[--C--] will be displayed while the reading is storedThe display will show [done] if the calibration was successful. Thebalance returns to the previous application mode and is ready for use. Internal Calibration (not available for SJX/E models)With balance turned on:Step-1 Press and hold Menu until [mMeNU] (Menu) is displayed.Step-2 Release the button, the display will show [C.A.L].Step-3 Press Yes to accept. [In.CAL] will then be shown.Step-4 Press Yes to begin the Internal calibration. Zero reading is storedautomatically.Step-5 Pull the InCal handle while the display show[PULL].Step-6 Push the InCal handle while the displayshow [PUSH].Step-7 Repeat the step-5, and step-6 one moretime.Step-8 The display will show [done] if the calibration was successful. Thebalance returns to the previous application mode and is ready for use.Visit www.balances.com for balances, scales & calibration weightsVisit www.scaleman.com to purchase Ohaus Scout SJX Scale