EN-8 OHAUS Valor TM 1000 Series4.3 CALIBRATIONSpan Calibration:Turn on the scale. When display flashes all segments ("888888") press and holdZERO and SELECT at the same time until "SET--- " is displayed at the end of poweron sequence. Press the SELECT key until "SET-F1" is shown on the display.Press ZERO key to display "A____x" (x=divisions in grams:1,2,5,10.20.50).Press ZERO key to accept the default division setting which is shown on the display.Do not change default.Display shows "b____x" (x=decimal point:0,1,2,3,4). Press ZERO key to accept thedefault decimal point setting which is shown on the display. Do not change default.Display shows "Cxxxxx" (xxxxx=Capacity of scale). Press ZERO key to accept thedefault Capacity setting which is shown on the display. Do not change default.Display shows "dx" (x=Zero tracking setting:0-9 from 0.125d to 1.25d). PressZERO key to accept the default setting which is shown on the display.Press ZERO and display will show "Load 0". Make sure pan is empty then pressZERO key to accept. Display will show "CAL 0" briefly, then "Load F” will bedisplayed.Press ZERO to display "Fxxxxx" (xxxxx=Span weight ). To accept default spanweight, put displayed weight on the scale, and press ZERO to accept.If span weight is not available, user can select a lower span calibration point value.Use the SELECT (scroll to next digit), TARE (decrease blinking digit) and M+(increase blinking digit) keys to increment or decrement span calibration values.Once desired span calibration weight is set, place displayed weight on the scale,and press ZERO to accept. After completion, scale will be in weighing mode.To abort calibration at any point, turn scale off. To restart calibration, followcalibration procedure from beginning.