50 Refer to the Solana assay package insert for result interpretation.Figure 56An amplification curve of the target nucleic acids for each test can also be viewed by an Administrator.Note: The curve is presented for information only and should not be used to override the result (Positive,Negative, or Invalid) generated by the software. From the Sample Results screen (Figure 56), touch the row with the relevant sample. The real-time curve of the selected tube is shown in a separate window (Figure 57). Move up and down the page using the scroll bar.Figure 57Managing ResultsRun data can be printed, exported to a USB flash drive, or deleted. Note: Official records of all test results aredisplayed and/or printed. Saved data (results) are for convenience only and are to be managed by theAdministrator.