Ranger Irrigation Fluid Warming System English11SpecificationsPhysical Characteristicswarming unit4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long;wt.: 7 lb. 7 oz. (3.4 kg)classificationClassified under IEC 601-1 Guidelines (and other national versions of the Guidelines) as Class I, Type B,Ordinary Equipment. Not suitable for use in the presence of flammable anesthetic mixtures with air orwith oxygen or nitrous oxide. Classified by Underwriters Laboratories Inc. with respect to electric shock,fire, and mechanical hazards only, in accordance with UL 60601-1 and in accordance with Canadian/CSAC22.2 No. 601.1.Electrical characteristics Temperature characteristicsinput voltage set point temperature100-120 or 220-240 VAC 41°Coperating frequency over-temperature alarm100-120 VAC, 50/60 Hz 48°C220-240 VAC, 50 Hzmaximum heating power under-temperature alarm900 W 33°Cfuse over-temperature cutoff10 Amp 50°CLeakage CurrentMeets leakage current requirements in accordance with UL 60601-1 and IEC 601-1.Environmental Conditionsoperating temperature range15° to 40°C (59° to 104°F)storage temperature range-20° to 45°C (-4° to 113°F)operating humidity10 to 85% RH, noncondensingatmospheric pressure range50 kPa to 106 kPa