User's Guide ______________________________________________________________________60 __________________________________________________________________ M211060EN-CPerformance (HMP110)Relative HumidityMeasurement range 0 ... 100 %RHAccuracy (incl. non-linearity, hysteresisand repeatability)temperature range +0 ... +40 °C0 ... 90 %RH ±1.7 %RH90 ... 100 %RH ±2.5 %RHtemperature range -40 ... 0 °Cand +40 ... +80 °C0 ... 90 %RH ±3.0 %RH90 ... 100 %RH ±4.0 %RHFactory calibration uncertainty (+20 °C) ±1.5 %RHHumidity sensor Vaisala HUMICAP® 180RTemperatureMeasurement range -40 ... +80 °CAccuracy over temperature range0 ... +40 °C ±0.2 °C-40 ... 0 °C, +40 ... +80 °C ±0.4 °CTemperature sensor Pt1000 RTD 1/3 Class B IEC 751DewpointMeasurement range -40 ... +80 °CAccuracy (incl. non-linearity,hysteresis and repeatability)temperature range 0 ... +40 °Cwhen temperature - dewpoint< 15 °C ±1 °C15 ... 25 °C ±2 °Ctemperature range -40 ... 0 °Cand +40 ... +80 °C whentemperature - dewpoint < 15 °C ±2 °CPerformance (HMP110T)TemperatureMeasurement range -40 ... +80 °CAccuracy over temperature range0 ... +40 °C ±0.2 °C-40 ... 0 °C, +40 ... +80 °C ±0.4 °CTemperature sensor Pt1000 RTD 1/3 Class B IEC 751