6. Troubleshooting6.1 Problem SituationsTable 9 Troubleshooting TableProblem Possible Cause SolutionDisplay shows one or moreerror codes.Various causes. Check meaning of the errorcode(s) and proceedaccordingly. See Error Codes(page 41).You are adding a new RFL100Data Logger to the system butit is not coming up as a NewDevice in viewLinc.RFL100 is not in range of anAP10 that has installationmode turned on, and capacityto add more data loggers(maximum 32 for each AP10).Turn on installation mode in anAP10 that is within 100 m of thedata logger, and has capacityto add more data loggers.RFL100 is connected to anAP10 that is not connected toviewLinc.Verify the following:• AP10 is connected to thenetwork and has an IPaddress.• AP10 is configured toconnect to the correctviewLinc Enterprise Server.• viewLinc connection of theAP10 is OK.RFL100 has been previouslyaccepted to a differentviewLinc system.Press the Release button of theRFL100 and verify that themessage NWK REL appears onscreen. The RFL100 is nowready to join any AP10 that is ininstallation mode.RFL100 has joined an AP10 buthas not been accepted to theviewLinc system. The AP10 isno longer in range, but RFL100cannot change to a differentaccess point as it is pending tobe accepted to the system.RFL100 User Guide M211861EN-A40