216.0 SpecificationsFreeChlorinePVCFreeChlorinePEEKChlorineDioxidePVCChlorineDioxidePEEKRange (W100,W600, W900)0.03-20 mg/l (reduce the range by a factor of 3 if using theoptional cleaning attachment)Range(WebMaster)0.03-8 mg/l (reduce the range by a factor of 3 if using theoptional cleaning attachment)Resolution 0.01 mg/lSensitivity NaOCl, Ca(OCl)2, Cl2.O3 (900%), ClO2 (400%),ClO2- detectedClO2.Cl2, ClO2- < 2% detectionFlow Rate ofSample20 – 100 liter/hour (constant, 80 l/hr. optimum if using theoptional cleaning attachment, 50 l/hr. without)pH range ofsamplepH 5.00 – 9.00 (pH must bestable within ±0.10)pH 1.00 – 12.00Conductivityrange of sampleNo LimitResponse time 30 secondsConditioningtime1 hour – 2 days depending on water qualityOperatingPressure0-8 bar (0-116 psi) – 104033 flow cell, 0-4 bar (0-58 psi) –192011 flow cellOperatingTemperature0 – 50 C 0 – 70 C 0 – 50 C 0 – 70 CStorage Frost-protected, dry and without electrolyte – no limitFlow cellconnectors6 mm ID x 8 mm OD tubing fittings in ¼” straight threadedholesElectrolyte P/N 104039, 50 mlPowerrequirements±5 VDC, 5 mASignal 0 to -2000 mV DCMaximum cablelength30 meters (100 feet)Extension cable 4 conductor 24 AWG shielded (Walchem 100084)Materials of constructionSensor PVC-U PEEK PVC-U PEEKFlow cell PMMA and Natural PVDF (104033) or PVC (192011)Mounting rings PETP, FKM, Silicone and Natural PVDF (104033) or PVC(192011)OptionalCleaning capPVDF, Acrylic (PMMA), Silicone, Ceramic ballsOptional flowswitchSwitch: Stainless steel, Polyester (not wetted)Float: PEEK, epoxy