Dental X-Ray Unit Z7070kVp - 7mAFuseFuse holderMobile Model Fig.02Dental X-Ray Unit Z7070kVp - 7mA5. Additional Information’sFollow the requerements from the table below, when the technician installing the Z70:5.2. Fuses Replacement16When necessary, the main power supply fuses must be changed, according the Fig.02, below.The fuse characteristics are described at the pag.04, item 2.1.Replacing the Main FusesElectrical Installation Requirements5.1.Additional Information’sElectrical Power RequirementsThe system requires a three-wire electrician-supplied powersupply. The three-wires provide two power lines (L) Line and(N) Neutral and a (G) Ground. Refer to template forrecommended AC main box loca on.* Informa on considered for 120V. For 230V, consult the XZeal.Line Voltage* 120 VAC ±10%Main Fuse 10 AMPCircuit Breaker 15 AMPWire length and gauge requirementsMaximum length of wire and minimum gauge wire (AWG) from thepower panel box to the Base unit.Accomplish with the local electrical regula ons law.Avoid malfunc on, risks and injuries, using an effec ve ground.0-50 feet 14 gauge minimum50-100 feet 12 gauge minimum100-150 feet 10 gauge minimum150-250 feet 8 gauge minimum250-400 feet 6 gauge minimum