Principles of Operation Section 8or less have larger electrodes that are closely-spaced.CHOOSING A CONDUCTIVITY CELLDecide which cell will be the most useful for your conditions by considering the conductivity ofthe solution you want to measure and the size of the sample.0 0.1 1.0 10 100 1000 μS/cm1 10 100 1000 mS/cm3440340134183256,3255, 34463254,3253,3252,ConductivityThe chart above reflects general guidelines. Refer to cell specifications for details.The multipoint calibration option on the YSI Model 3200 Conductivity Instrument gives greaterflexibility and performance when using one cell over a wide conductivity range. This featureallows users to calibrate one cell with up to 5 different conductivity standard solutions. Whenanalyzing a sample, the instrument selects the calibration value that is closest in conductivity tothe sample.3200 S ERIES CONDUCTIVITY CELLSDip cells are generally used for routine conductivity measurements. The 3254 Fill Cell isdesigned for small sample or high throughput work. It requires only 5 ml of sample and cantake measurements quickly from one sample to the next.The 3255 Flow Cell is designed for in-line conductivity measurements, such as forultrapure water systems.YSI Incorporated Model 3200 43