ZOLL M Series Invasive Blood Pressure 13Printing IBP DataSee the M Series Operator’s Guide Recorder Operation section for instructions on how to print data. Ifyou are currently displaying IBP traces, press the RECORDER button to print a strip chart that includesthe following values for the currently displayed trace:• One or two IBP traces (whichever is currently displayed).• Systolic blood pressure.• Diastolic blood pressure.• Mean pressure.• Time of measurement.Maintaining the IBP OptionSee the M Series Service Manual for instructions on maintaining the IBP option.Performing the Daily Checkout ProcedureThe following is a checkout procedure that should be performed daily to ensure that the IBP option isfunctioning properly. In addition, this daily checkout procedure ensures that medical personnelmaintain familiarity with the proper use of the IBP option.Use only 5uV/V/mmHg IBP simulators. See "Ordering Accessories" on page 16 for further details. Formore specific instructions on how to accomplish each step listed below, refer to the related section inthis manual.To perform the daily checkout procedure:1. Connect the IBP simulator cable to the M Series unit and the IBP simulator.2. Set the simulator to zero.3. Zero the transducer.4. Set the simulator to simulate 50 mmHg.5. Verify that the M Series displays systolic, diastolic and mean pressures of 50 mmHg +/- 6mmHg.6. Set the simulator to simulate 250 mmHg.7. Verify that the M Series displays systolic, diastolic and mean pressures of 250 mmHg +/- 6mmHg.