General InformationStratus OCT Model 3000 Service Manual 3 2660021155438 ASafety InformationWARNING: To prevent electric shock, the instrument must be pluggedinto an earth grounded outlet. Do not remove or disable the groundpin. Only an authorized Carl Zeiss Meditec service representative mayinstall the instrument.WARNING: Do not use the printer or the instrument with an extensioncord or a power strip (multiple portable socket outlet). Failure toobserve this warning could result in electrical shock to the patientand/or examiner.WARNING: Do not open the instrument covers. Opening the instrumentcovers could expose you to electrical and optical hazards.WARNING: If the instrument is externally connected to non-medicalperipheral devices (i.e. printer, storage devices, etc.), the completesystem must comply with the system requirements in standard IEC60601-1. This standard requires the usage of an Isolation Transformerto power the non-medical peripheral device(s) if located within 1.5 mfrom the patient. If the peripheral device is located outside the patientenvironment (beyond 1.5 m) and is connected to the Stratus OCT, aseparation device must be used or there shall be no metal to metalconnection between the non-medical peripheral device and the StratusOCT.The person or the responsible organization connecting additionaldevices or reconfiguring the system must evaluate the complete systemto ensure compliance to the applicable IEC 60601-1 requirements.The instrument operator must not attempt to touch the patient and theperipheral device simultaneously.WARNING: Do not reconfigure system components on the table, noradd non-system devices or components to the table, nor replaceoriginal system components with substitutes not approved by Carl ZeissMeditec. Such actions could result in failure of the table heightadjustment mechanism, instability of the table, tipping and damage tothe instrument, and injury to operator and patient.WARNING: This instrument may cause ignition of flammable gases orvapors. Do NOT use in the presence of flammable anesthetics such asnitrous oxide, or in the presence of pure oxygen.WARNING: Avoid tipping. Do not use the instrument on an uneven orsloped surface. Also, do not roll the table in deep pile carpet or overobjects on the floor such as power cords. Failure to observe theseprecautions could result in tipping of the instrument and/or table andresulting injury to operator or patient and damage to the instrument.CAUTION: Federal law restricts this device to sale by or on the order of aPhysician or Practitioner.