Operating Instructions8.2 Performing Cold Treatment 8Seite 21Switch on the device Switch the device on using the main switch (10); the display will illuminate.Initialization During initialization, the current status can be seen on the display(self-test, pre-cooling).MenüOK Start/ StopVorkühlung0% 100%Notice: Depending on the preset settings, Cryo 6 Physio or Cryo 6 Derma, the start screenor start sequence is represented differently as a factory default.Please see the next page for illustrations of the two start screens.SelectionProgram and therapytimeThe device is operationally ready as soon as the therapy screen (here "Physio")appears.OK Start/ Stop10:00P2 7P PUsing the program selection keys (14 and 15), set the desired treatment programand using the timer keys (16 and 17), set the desired therapy time.Notice: Prior to starting treatment, the fan level cannot be changed.Start of the treatment The program is started by pressing the Start/Stop key (18).During treatment, the keys can be used to change fan level (14 and 15) and therapytime (16 and 17).7 05:30OK Start/ StopNotice: If parameters are changed during treatment, after the treatment, the system willrestore the factory default settings or, if a favorite program is stored, the favoriteprogram settings will be restored.End of the treatment An acoustic signal indicates the end of treatment and the fan is automatically shutoff. This also applies when prematurely aborting treatment using the “Start / Stop”key (18).